1. For any sensor enter the following service terminal command:
LASTVAL <measurement_name> status
For example:
/ > LASTVAL TAMeasQMH101_1 status Status:1 Value:0
For details on the LASTVAL command, refer to section
LASTVAL Command on page 198.
2. For conventional sensors, open the Sensor Calibration tab in
MAWS Terminal, see Figure 116 on page 129. The list of the
MAWS sensors is displayed in the tab. The Status column shows
the sensor status. To read the latest sensor statuses and
measurement values, click the Refresh button. Table 15 on page
130 describes the other columns in the Sensor Calibration tab.
0312-038
Figure 149 MAWS Station Settings Window: Sensor Calibration
Tab
where
Status:1 = The Status field in the output is the value status.
Valu e: 0 = Th e Value field shows the sensor status value.
CAUTION
When you upgrade a previous MAWS version to 3.06 or above, the
following sensors will not operate correctly in the Sensor Calibration
tab until they are removed from the setup and then re-created: CM6B,
CM11, QMS101, QMS102, and QMN101.