Chapter 6 ______________________________________________________________ Maintenance
VAISALA______________________________________________________________________ 173
Changing the HUMICAP®180 Humidity Sensor
Unscrew the filter. Remove the damaged sensor and mount a new
HUMICAP®180 humidity sensor in its place. Handle the sensor with
care. Calibrate the probe using a two-point calibration procedure.
Humidity Calibration
For a high-accuracy two-point calibration, use a Vaisala HMK15
calibrator and saturated salt solutions. Refer to the appropriate manual
for details.
Easiest way of doing the calibration is to use HMI41 Indicator with the
HMH40-MAWS Handle for the probe head. Note that this setup is only
for relative humidity and no temperature readings are displayed.
0406-072
Figure 145 HMH40-MAWS Handle
Leave the calibrator and the probe head in the same space for at least
four hours so that their temperatures have time to equalize. Unscrew the
plastic grid of the probe.
The calibration is done first for the dry end (<50 %RH) and then for the
wet end (>50 %RH) by adjusting the trimmer potentiometers marked
with D (dry) and W (wet). The potentiometers are located under a
protective plug. Use a ceramic with 2.5 mm (0.1 in.) blade for adjusting
the potentiometers.
NOTE
If zero point is calibrated in nitrogen (N2), the minimum output signal
of 0.008 V corresponds to a relative humidity of 0.8 %RH.