Chapter 4 _________________________________________________________________ Operation
VAISALA________________________________________________________________________ 87
Example:
0 ... 5 V voltage output selected for channel 1 and 4 ... 20 mA selected
for channel 2.
OFF ON Selection
1
2
Voltage output selecte
d
3
4
5
6
0 ... 5 V selecte
d
7
8
1
Current output selecte
d
2
3
4
4 ... 20 mA selecte
d
5
6
7
8
NOTE
If you have customized the error output setting (AERR), check that
the set error values are still valid after changing the output
mode/range, see section Analog Output Fault Indication Setting on
page 90.
Analog Output Quantities
NOTE
The relative humidity and temperature readings are not those of the
process itself but the ones measured after the cooling process.
Therefore, they should not be selected as output quantities. They are
meant to be used in calibration or during installation.
Use the display/keypad to change and scale the analog output
quantities.