User's Guide _______________________________________________________________________
100 __________________________________________________________________ M210762EN-A
CLOSE
The CLOSE command switches the transmitter back to the POLL
mode.
Example:
>OPEN 2 (opens the line to transmitter 2, other
transmitters stay in POLL mode)
>CRH(for example, calibration performed)
...
>CLOSE (line closed)
SDELAY
With the sdelay command you can set delay (response time) for user
port (RS-232 or RS-485), or view currently set delay value. Value
corresponds to tens of milliseconds (eg. 5 = 0.050s minimum answer
delay). The value can be set between 0...254.
Example:
>sdelay
Serial delay : 0 ? 10
>sdelay
Serial delay : 10 ?
>
Sensor Functions
AutoCal
To obtain the best possible accuracy in measurements taken in dry
environments, DMT345 and DMT346 have a built-in AutoCal feature.
During the AutoCal, the transmitter adjusts the dry-end reading to
correspond to the calibrated values. This is a unique and patented
method to avoid errors in accuracy when monitoring low dewpoints.