Chapter 6 _______________________________________________________________ Maintenance
VAISALA_______________________________________________________________________ 123
Error Message Action
probe and the probe cable. Clean
the probe from dirt water, ice or
other contaminants.
Temperature measurement
malfunction
Check the integrity of the humidity
probe and the probe cable. Clean
the probe from dirt water, ice or
other contaminants.
Temperature sensor current leak. Check the integrity of the humidity
probe and the probe cables. Clean
the probes from dirt, water, ice or
other contaminants.
Internal ADC read error Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Internal EEPROM read error Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Internal EEPROM write error Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Add-on module 1 (or 2) connection
failure
Turn off the power and check the
module connection. Turn on the
power.
Device internal temperature out of
range
Ensure that the operating
temperature is within the valid range.
Operating voltage out of range Ensure that the operating voltage is
within the valid range.
Internal analog voltage out of range Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Internal system voltage out of range Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Internal ADC reference voltage out
of range
Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Internal analog output reference
voltage out of range
Internal transmitter failure. Remove
the transmitter and return the faulty
unit to Vaisala Service.
Configuration switches for analog
output 1/2/3 set incorrectly
Check and re-set the switches, see
page
51.
EEPROM failure on add-on module
1 (or 2)
Disconnect the power and check the
analog output module connection.
Communication module installed in
incorrect add-on module slot
Disconnect the power and change
the communication module to
another module slot.
Unknown/incompatible module
installed in add-on module slot
1(or2)
Ensure that the module is
compatible with the DMT345/346.
You can also check the error message via the serial interface by using
the command ERRS. In case of constant error, please contact the
Vaisala Service Centers (contact information on page 125).