User's Guide _______________________________________________________________________
30 ___________________________________________________________________ M210762EN-A
The following numbers refer to Figure 16 on page 29:
1 = Process or duct wall
2 = Flange
3 = Sealing ring
4 = Probe
NOTE
When the temperature difference between the process or duct and the
surroundings is large, the probe must be installed as deep in the
process or duct as possible. This prevents errors caused by heat
conduction in the probe cable.
DMT346 Probe and Cooling Set Mounting
22.5
ø1
334.5
ø13.5
357
1 2
0604-020
Figure 17 DMT346 Probe Dimensions in mm
The following numbers refer to Figure 17 above:
1 = Stainless steel grid
2 = Cable (length 2 m, 5 m or 10 m)
Kommentare zu diesen Handbüchern