User's Guide _______________________________________________________________________
114 __________________________________________________________________ M210762EN-A
0605-019
Figure 51 Proceeding with T
d/f
Adjustment
5. Enter the actual reference frostpoint temperature.
6. The transmitter performs T
d/f
adjustment. This may take up to
six (6) minutes.
0605-020
Figure 52 Completing T
d/f
Adjustment
7. The adjustment is now completed.
NOTE
Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.
One-point Dewpoint Adjustment
Using Serial Line
1. Let the sensor stabilize for at least one (1) hour.
2. Press the ADJ button on the motherboard to enable adjustments.
3. Use the AØ1 command to perform the adjustment.
Syntax: AØ1