1. Make a round 89.5 + 0.5 mm hole on the process wall.
2. Weld the tube of the mounting flange tightly on the inner metal
plate of the process wall. If the process wall is more than 125
mm thick, a lengthening piece (max. 50 mm) can be welded on
the mounting tube (for walls thicker than 175 mm, see Figure 23
on page 35).
3. The cooling set is mounted in a vertical position. Tighten the
screws properly in order to ensure a thermal contact.
4. Unscrew the locking screws on the cooling bar so that you can
push the sensor head into the bar.
NOTE
Push the sensor head deep enough: make sure that the marking hole
meets the end of the bar.
5. Lock the sensor head in place by tightening the locking screws
on the bar.
Probe and Cooling Set Mounting with Process Shut Down
- Complete the entire mounting process as instructed on page 33.
Probe and Cooling Set Mounting with Process Running
- First mount the cooling element (and fins) to warm it up; tightly
plug the hole for the probe in the cooling element for the duration
of the warmup period.
- After a few hours, complete the installation by installing the probe.
This procedure should lessen the amount of condensation that builds
up in the hole.