Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 89
Table 26 Pressure commands
Command Description
PRES [hPa] Set the value for pressure compensations
XPRES [hPa] Set the value for pressure compensations,
temporarily
PFIX Select pressure compensation using either a
fixed value or using measured value
AVRG Set pressure average period
HHCP Set height offset for HCP calculation
HQNH Set height offset for QNH calculation
HQFE Set height offset for QFE calculation
PSTAB Set the pressure stability indicator
PDMAX Set the pressure difference limit
Table 27 GPS Commands
Command Description
0100P9 Data output query
0200P9 Data output query
9900P9 Data output query
9900SN Serial number query
Table 28 Other Commands
Command Description
? Output information about the device
?? Output information about the device in POLL
mode
ECHO [ON/OFF] Turn the serial interface echo ON/OFF
ERRS List present transmitter errors
FILT Set the result filtering
FIND All devices in POLL mode send their addresses
HELP List the most common commands
LOCK Lock the menu and disable the keypad
VERS Display the software version information
XHEAT Sensor heating
Getting Measurement Message from
Serial Line
Starting Continuous Outputting
R
Enter the R command to start the continuous output of measurements.
Example:
>r
P= 1021.6 hPa T= 23.2 'C RH= 5.8 %RH
>