Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 115
Example: 0 ... 5 V voltage output selected for channel 1 and 4...20 mA
selected for channel 2.
OFF ON Selection
1
2
Volta
e out
ut selecte
3
4
5
6
0 ... 5 V selecte
7
8
1
Current out
ut selecte
2
3
4
4 ... 20 mA selecte
5
6
7
8
NOTE
If you have customized the error output setting (AERR), check that the
set error values are still valid after changing the output mode/range, see
section Analog Output Fault Indication Setting on page 118.
Analog Output Quantities
NOTE
For best accuracy, pressure must always be output using Ch3, if
available.
Use the display/keypad to change and scale the analog output quantities.
1. Press any of the arrow buttons to open the MAIN MENU.
2. Select Interfaces and press the right arrow button.
3. Select Analog outputs and press the right arrow button.
4. Select Output 1/2/3 and press the right arrow button.
5. Select Quantity and press the CHANGE key.
6. Select the quantity by using the up/down arrow buttons. Press the
SELECT key to confirm your selection.