USER'S GUIDE____________________________________________________________________
126 _________________________________________________________________ M210796EN-E
Networking Commands
Set the RS-422/485 interface by using the following commands. The
other serial line commands are presented in section List of Serial
Commands on page 87.
RS-485 configuration commands SERI, ECHO, SMODE, INTV, and
ADDR may be entered by using either the service port or RS-422/485
port. Also the optional display/keypad can be used, see section User Port
Serial Settings on page 102.
SDELAY
With the SDELAY command you can set delay (response time) for user
port (RS232 or RS485), or view currently set delay value. Value
corresponds to tens of milliseconds (eg. 5 = 0.050s minimum answer
delay). The value can be set between 0 ... 254.
Example:
>sdelay
Serial delay : 0 ? 10
>sdelay
Serial delay : 10 ?
SERI
Use the SERI command to input RS-485 bus settings.
SERI [b p d s]
where
b
= bit rate (300, 600, 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200)
p
= parity (n = none, e = even, o = odd)
d
= data bits (7 or 8)
s
= stop bits (1 or 2)
Seitenansicht 127