USER'S GUIDE____________________________________________________________________
144 _________________________________________________________________ M210796EN-E
1-point Adjustment Using Serial Line
NOTE
Making adjustments is possible only after adjustments are unlocked. To
unlock the adjustment menu, press the ADJ button on the motherboard
of the transmitter.
LCI
Use the command LCI
- to activate or deactivate the linear adjustment function
- to enter new linear offset and offset/gain pressure corrections to the
transmitter
- to edit existing linear offset and offset/gain pressure corrections.
Key in the linear corrections for each pressure transducer separately.
NOTE
The new linear corrections will always cancel the previous corrections as
well as the valid date of calibration of the transmitter.
Example:
>lci
P1 linear adj. : OFF ? on
P1 1.reading : 0.000 ? 950
P1 1.correction: 0.000 ? 0.22
P1 2.reading : 0.000 ? 1120
P1 2.correction: 0.000 ? 0.15
LC
Use the command LC to view current status of the linear offset and
offset/gain pressure corrections.
Example:
>lc
P1 linear adj. : ON
P1 1.reading : 950.000
P1 1.correction: 0.220
P1 2.reading : 1120.000
P1 2.correction: 0.150
MPCI
Use the command MPCI
- to activate or deactivate multipoint adjustment function