Vaisala PTU300 Betriebsanweisung Seite 107

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 186
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 106
Chapter 4 ________________________________________________________________ Operation
VAISALA ______________________________________________________________________ 105
Pressure Average Calculation
Pressure
The averaging data filter calculates an average pressure over a certain
period of time. The lowest measurement noise is achieved with the
extended filtering. There are three filtering levels available.
AVRG [x] <cr>
where
X = 1 ... 60 s (default: 1 s)
.
The AVRG command is used to set and inspect the averaging time
during which the individual measurement samples are integrated to get
an averaged reading. The averaging time is the total averaging time of the
transmitter.
Note that if the averaging time is defined to be long, the settling time at
power-up will be long, too.
A minimum of one-second averaging time is recommended per each
pressure transducer. These selections are used as the factory setting
averaging times.
Example of setting the averaging time to 60 seconds (WMO averaging
time for barometric pressure measurement):
>avrg <cr>
P1 average : 1 s ? 60 <cr>
>avrg <cr>
P1 average : 60 s ? <cr>
Relative Humidity (RH) and Temperature (T) Filtering
Table 32 Filtering Levels for Relative Humidity (RH) and
Temperature (T)
Setting Filtering level
OFF No filtering
ON (default) Standard = short filtering (approximately 15 s moving
average)
EXTENDED Extended filtering (default: approximately 1 min
average)
Use display/keypad to set the filtering level.
Seitenansicht 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 185 186

Kommentare zu diesen Handbüchern

Keine Kommentare