Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 49
Adjustment Procedure (HMP110)
NOTE
If you are adjusting a HMP110 or HMP110T probe that is in the analog
output mode, see section Terminal Application Settings for Probes in
Analog Mode on page 30 for instructions on how to start the probe in
RS-485 mode.
One-Point Adjustment of RH
Measurement (HMP110)
To perform a one-point adjustment to the capacitance measurement of
the HMP110 you need:
- The Vaisala USB cable (Vaisala order code: 219690)
- PC with a terminal application.
- One humidity reference. One-point calibration with a single
< 50 %RH reference will adjust the offset of the measurement. One-
point calibration with a single > 50 %RH reference will adjust the gain
of the measurement.
The procedure below uses the HMK15 Humidity Calibrator. LiCl salt
(11 %RH) is used as the reference point.
1. Connect the USB cable to the PC, but do not connect it to the probe
yet.
2. Insert the probe in the LiCl salt chamber of the humidity calibrator
(11 %RH).
3. Start a terminal application and set the correct connection settings.
The default serial settings are 19200 8 N 1. Remember to check
which COM port the USB cable is using.
4. Wait for 20 – 40 minutes for the humidity to stabilize.
5. Start the terminal session and connect the USB cable to the probe.
If your probe is in the analog output mode, you need to press Enter
a few times to start it in RS-485 mode.
6. Verify that the connection works by giving the ? command:
?
If the probe does not respond with device information:
- Disconnect the USB cable from the probe and retry. If your
probe is in the analog output mode, press Enter a few times
immediately after connecting the USB cable.
- Try the ?? command in case the probe is in POLL mode, open
the line using OPEN command if necessary.
- Check your serial line settings and cable connections.