Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 47
CHAPTER 5
MAINTENANCE
This chapter provides information that is needed in basic maintenance of
the product.
Periodic Maintenance
The humidity measurement accuracy of the HMP60 and HMP110 should
be calibrated yearly. When calibration indicates that accuracy is not
within specification:
- HMP60: change the INTERCAP® Sensor.
- HMP110: adjust the measurement yourself, or have it adjusted at a
Vaisala Service Center.
Light cleaning of the probe, and replacement of the filter should be done
only when necessary.
Cleaning
The probe body can be wiped clean with a soft, lint-free cloth moistened
with mild detergent. Do not use solvents or compressed air.
Note that wiping the membrane filter or stainless steel sintered filter may
block its pores and/or deposit residue on the filter. If the filter is heavily
contaminated, replace it.