List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
Table 8 List of Serial Commands
Command Description
? Output information about the device
ADJD View the adjustment date
AMODE View or set the analog output mode
AOVER [ON/OFF] Allow analog outputs to exceed their range
10%
CDATE View the calibration date
CODE View the order code of the probe
CRH Calibrate and adjust RH measurement
CRHCLR Clear adjustment of RH measurement
CT Calibrate and adjust T measurement
CTCLR Clear adjustment of T measurement
CTEXT View the calibration information field
ERRS List present probe errors
FILT [0.001 … 1] Set the result filtering
FRESTORE Restore factory settings
HELP List available commands
INTV [0 ... 255 S/MIN/H] Set the continuous output interval (for RUN
mode)
L Displays user adjustment parameters
R Start the continuous outputting
RESET Reset the probe
S Stop the continuous outputting
SDELAY [0 … 255] View or set serial line answer minimum
delay
SEND [0 ... 99] Output the reading once
SNUM View the serial number of the probe
UNIT Select metric or non-metric output units
VERS View software version of the probe