26 CONTENTS
Setting of Measurement Parameters
Before using the barometer you have to set up the parameters. The following
commands will give you an overview in terms of possible measurement settings.
To check the actual settings use ?<cr>.
To set the number of measurements performed in one minute use the com-
mand:
.MPM.xxxx<cr>
xxxx = number of measurements per minute (6...4200)
To set the number of measurement points that are used for calculating
an average pressure reading use the command:
.AVRG.xxx<cr>
xxxx = number of measurements (not a moving number) of which the
average is calculated (0...255)
NOTE: Make sure to reset the barometer with the command .RESET<cr>
to activate the new settings. The settings can be checked with the
command ?<cr>.