Vaisala HUMICAP MMT310 Wartungshandbuch Seite 13

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 123
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 12
22 23
The HMP365 probe is designed for high
temperature environments.
The HMP367 probe is constructed to
be installed in environments with high
humidities.
The HMP368 probe enables flexible
installation in pressurized pipelines.
Dimensions
Dimensions in mm (inches)
Dimensions
Dimensions in mm (inches)
Dimensions
Dimensions in mm (inches)
Technical Data
HMP365 for high temperature
Temperature range -40 ... +180 °C
(-40 ... +356 °F)
Probe cable length 2, 5 or 10 meters
Probe diameter 13.5 mm
Installation
Mounting flange 210696
Cable Gland M20x1.5
with splitting seal HMP247CG
Technical Data
HMP367 for high humidities
Temperature range -40 ... +180 °C
(-40 ... +356 °F)
Probe cable length 2, 5 or 10 meters
Probe diameter 12 mm
Installation
Duct installation kit 210697
Cable Gland M20x1.5
with splitting seal HMP247CG
Swagelok for 12mm
probe, 3/8" ISO thread SWG12ISO38
Swagelok for 12mm
probe, 1/2" NPT thread SWG12NPT12
Technical Data
HMP368 for pressurized
pipelines
Temperature range -40 ... +180 °C
(-40 ... +356 °F)
Pressure range 0 ... 4 MPa
Probe cable length 2, 5 or 10 meters
Probe diameter 13.5 mm/12 mm
Two probe lenghts available.
Installation
Fitting body ISO1/2 solid structure
DRW212076SP
Fitting body NPT1/2 solid structure
NPTFITBODASP
Ball valve ISO 1/2 with welding joint
BALLVALVE-1
Length for standard / optional probes
*freely user adjustable length
R ½ ISO 7/1 or NPT½"
41 - 149/371 (1.61 - 5.87/14.6)*
194/416 (7.64/16.4)
232/454 (9.13/17.9)
14 (0.55)
Ø13.5 (0.53)
41 (1.61)
41 (1.6)
51 (2.0)
ø13.5 (0.53)
192 (7.56)
253 (9.96)
ø 12 (0.47)
37.5 (1.48)
79.5 (3.13)
99.5 (3.92)
M9×1
drilling 16...22 (0.63...0.87)
50 (1.97)
75 (2.96)
Installation flange:
stainless steel
Technical data
Performance
RELATIVE HUMIDITY
Measurement range 0 ... 100 %RH
Accuracy (including non-linearity, hysteresis, and repeatability)
with Vaisala HUMICAP
®
180 or 180R for typical applications
at +15 ... +25 °C (59 ... +77 °F) ± 1.0 %RH (0 ... 90 %RH)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C (-4 ... +104 °F)
±(1.0 + 0.008 x reading)
%RH
at -40 ... +180 °C (-40 ... +356 °F)
± (1.5 +0.015 x reading)
%RH
with Vaisala HUMICAP
®
180
L
2 for application with demanding
chemical environment
at -10... +40 °C (14 ... +104 °F) ± (1.0 + 0.01 x reading) %RH
at -40 ... +180 °C (-40 ... +356 °F) ± (1.5 + 0.02 x reading) %RH
Factory calibration uncertainty (+20 °C) ± 0.6 %RH (0 ... 40 %RH)
± 1.0 %RH (40 ... 97 %RH)
(Defined as ±2 standard deviation limits. Small variations
possible, see also calibration certificate.)
Response time (90 %) at +20 °C (+68 °F) in still air
with grid filter 8 s / 17 s*
with grid + steel netting filter 20 s / 50 s*
with sintered filter 40 s / 60 s*
* with HUMICAP
®
180R sensor
TEMPERATURE
Measurement range -40 ... +180 °C (-40 ... +356 °F)
(depends on selected probe)
Typical accuracy of electronics at +20 °C (+68 °F) ±0.2 °C (0.36 °F)
Typical temparature dependence
of electronics 0.005 °C/°C (0.005 °F/°F)
Sensor Pt 1000 RTD 1/3 Class B IEC 751
Accuracy over temperature range
OTHER VARIABLES
Optionally available dewpoint temperature, mixing ratio,
absolute humidity, wet bulb temperature.
Operating environment
Temperature range
operating temp. range for
electronics
-40 ... +60 °C (-40 ... +140 °F)
with display -20 ... +60 °C (-4 ... +140 °F)
storage -40 ... +70 °C (-40 ... +158 °F)
Pressure range see probe specifications
Complies with EMC standard EN61326-1, Electrical equipment for
measurement, control and laboratory use - EMC requirements;
Industrial Environment.
NOTE! IEC 1000-4-5 complies only when using external EXi
approved surge arrester in the safe area.
°C
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
Inputs and outputs
Operating voltage 12 ... 28 V
with serial port (service mode) 15 ... 28 V
Analog outputs two-wire 4 ... 20 mA, one standard, one optional
Typical accuracy of analog outputs at +20 °C ±0.05% full scale
Typical temperature dependence
of analog outputs 0.005% / °C (0.005% / °F) full scale
Analog outputs connection via safety barriers
RS232C serial output for service use connector type RJ45
Display two-line LCD
Mechanics
Connections screw terminals, 0.33...2.0 mm
2 wires (AWG 14-22)
Cable bushings For 7.5...12mm or 10...15mm cable
diameters (M20)
Conduit fitting NPT 1/2" (M20)
Housing material G-AlS
i
10Mg (DIN 1725)
Housing classification IP66 (NEMA 4X)
Housing weight 950 g
Options and accessories
Duct installation kit (for HMP363/367) 210697
Mounting flange (for HMP365) 210696
Ball valve ISO 1/2 with welding joint
(for HMP368)
BALLVALVE-1
pressure range at +20 °C (+68 °F): 0 ... 20 bar (0 ... 290 psia)
(during installation max. 10 bar (145 psia)
Calibration adapter for HMK15 211302
Serial interface cable for PC
connectors RJ45 - D9 female 25905ZZ
Galvanic isolator 212483
Zener barrier (USA & Canada) 210664
Protection cover (for use in the
presence of combustible dust, ATEX)
214101
II 1 D IP65 T = 70 °C
Seitenansicht 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 ... 122 123

Kommentare zu diesen Handbüchern