User's Guide ______________________________________________________________________
Configuration Mode/
Measurement Mode and
Profile
Command
Description
Measurement mode
WS425 A/B NMEA
Extended profile
$WIP
Polls data from WMT700.
Measurement mode
WMT700 NMEA MWV
profile
$aabbQ,MWV
Polls NMEA MWV message from WMT700.
Measurement mode
WS425 A/B ASCII profile
I
Requests identification information from WMT700.
Measurement mode
WS425 A/B ASCII profile
Wx
Starts measurement based on averaging time and
fetches the data.
Measurement mode
WS425 A/B WAT11 profile
<esc><id>
Polls data from WMT700.
Measurement mode
SDI-12 profile
?!
Queries WMT700 for its address.
Measurement mode
SDI-12 profile
a!
Ensures that WMT700 is responding.
Measurement mode
SDI-12 profile
aI!
Queries WMT700 for its SDI-12 compatibility level,
model number, and firmware version number.
Measurement mode
SDI-12 profile
aAb!
Changes WMT700 address.
Measurement mode
SDI-12 profile
aC!
Starts concurrent measurement.
Measurement mode
SDI-12 profile
aCC!
Starts concurrent measurement.
CRC is included in the response.
Measurement mode
SDI-12 profile
aD0!
Fetches instant data from WMT700.
Measurement mode
SDI-12 profile
aM!
Starts measurement.
Measurement mode
SDI-12 profile
aMC!
Starts measurement.
CRC is included in the response.
Measurement mode
SDI-12 profile
aV!
Starts verification.