User's Guide _______________________________________________________________________
42 ___________________________________________________________________ M211060EN-F
Set Probe Address
Use the ADDR command to view or set the probe address. To operate in
the POLL mode, the probe must have an address. If multiple probes share
the same serial line, each probe must have a different address.
For a description of the serial interface modes, see section Set Serial Line
Settings on page 39.
ADDR [nn]<cr>
Example:
addr
Address : 0
Set Serial Interface Delay
Use the SDELAY command to view or set the serial interface answer
minimum delay.
SDELAY [delay]<cr>
Range 0 ... 255. Value corresponds to four
milliseconds (for example, 5 = 0.020 second
minimum answer delay)
Example:
sdelay
Serial delay : 30
sdelay 50
Serial delay : 50