Chapter 2 ___________________________________________________________ Product Overview
VAISALA ________________________________________________________________________ 17
Duct Installation Kit for HMP60, HMP110,
HMP110D and HMP110T
The duct installation kit includes a plastic pipe with a flange (Vaisala
order code: 215619). To install the probe with the duct installation kit,
drill a hole to the duct wall, assemble the probe to the duct installation
kit, slide the probe head through the hole, and attach the flange to the
duct wall with four screws. See page 22 for details.
1301-002
Figure 10 Probe Installation with the Duct Installation Kit
The following explanations refer to Figure 10 above:
1
=
Tension screw
Distance L can be adjusted and locked in place with the tension screw.