Vaisala DRYCAP DMT345 Betriebsanweisung Seite 137

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 157
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 136
Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA ______________________________________________________________________ 137
0706-044
Figure 70 Proceeding with T
d/f
Adjustment
5. Enter the actual reference frostpoint temperature.
6. The transmitter performs T
d/f
adjustment. This may take up to six
(6) minutes.
0706-045
Figure 71 Completing T
d/f
Adjustment
7. The adjustment is now completed.
NOTE
Several AutoCals can be necessary after this adjustment has been
performed until the transmitter reaches full accuracy.
One-point Dewpoint Adjustment Using
Serial Line
1. Let the sensor stabilize for at least one (1) hour.
2. Press the ADJ button on the motherboard to enable adjustments.
3. Use the AØ1 command to perform the adjustment.
Syntax: AØ1
Seitenansicht 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 156 157

Kommentare zu diesen Handbüchern

Keine Kommentare