Vaisala HUMICAP HMP46 Betriebsanweisung Seite 47

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 46
HMI41 AND PROBES
M210316EN-A Operating Manual
43
11.3 HMP46 probe
11.3.1 Relative humidity
Measurement range 0...100 %RH non-condensing
Accuracy (at +20 °C); maximum achievable accuracy
when calibrated against high quality,
certified humidity standards ±1 %RH (0...90 %RH)
±2 %RH (90...100 %RH)
when calibrated against
salt solutions (ASTM E104-85): ±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)
Temperature dependence:
-40
0
40
16080
120
2
1
0
-1
-2
-20 20 60
100
140
180
1.5
0.5
-0.5
-1.5
error %RH
T/°C
for temporary
use only
Typical long-term stability better than 1 %RH/ year
Response time (90%) at 20 °C
in still air with sintered filter 15 s
Humidity sensor HUMICAP
®
180
11.3.2 Temperature
Measurement range (sensor head) -40...+100 °C (temporarily +180 °C)
Temperature sensor Pt100 IEC 751 1/3 Class B
Accuracy at +20°C ±0.2 °C
Accuracy over the whole temperature range:
°C
-40 -20 0 20 40 60 80 100 120 140 160
T/°C
0.6
0.4
0.2
-0.2
-0.4
-0.6
0
180
Temperature dependence
of electronics 0.005 °C/ °C
Seitenansicht 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 71 72

Kommentare zu diesen Handbüchern

Keine Kommentare