HMD70U/Y Transmitters
U303en-1.1 Operating Manual
1997-07-31 11
6.3 General
Supply voltage range depends on the selected output signal.
When AC supply is used, an isolated source is recommended.
The current outputs require an optional output module,
part no.18945HM.
DC AC
0...1 V 10...35 V 9...24 V
0...5 V 14...35 V 12...24 V
0...10 V 19...35 V 16...24 V
0...20 mA (R
L
= 0 Ω) 10...35 V 11...24 V
0...20 mA (R
L
= 500 Ω) 20...35 V 17...24 V
Factory setting 0...1V. Other outputs selectable by jumper connections.
An output change causes an error which is less than 0.5 %RH
without recalibration.
Operating temperature range:
electronics -5...+55 °C
sensor head -40...+80 °C
Storage temperature range -40...+80 °C
Housing:
sensor head stainless steel
electronics housing cast aluminium
Cable glands
bushing for 7...10 mm (PG9) cable
(housing IP65/ NEMA 4),
part no 18941HM
or armoured cable glands part no 10528HM
Sensor protection:
standard membrane filter
(part no. 17039HM)
option stainless steel sintered filter
(part no. HM46670)
Connections screw terminals 0.5...1.5 mm