USER'S GUIDE____________________________________________________________________
4. From the list of parameters, select T
I/II
5. At this point you need to select the adjustment method. If the
reference probe is connected, two options are displayed: To
same as T
I/II
, 1-point adjustment or 2-point adjustment.
6. You are prompted to check the settings, make sure that the
pressure units of DM70 and DMT340 are the same. If not,
change the unit by pressing Unit on the corresponding row.
To Same as T
I/II
NOTE
This adjustment method automatically uses the reading of the
reference probe as the reference value for making the adjustment.
7. After selecting To same as T
I/II
, confirm the adjustment by
pressing Yes.
If you press No, a note is displayed on the screen reminding
you to follow the adjustment procedure described in the User's
Guide. Press OK and the adjustment mode display returns
and no changes are made.
1-Point Adjustment
7.
After selecting 1-point adjustment, press Select to
continue.
8. Set the probe to the reference temperature. You can follow the
stabilization from the graph display by pressing
Graph.
When the reading has stabilized, press
Ready.
9. Use the arrow buttons to give the reference temperature value
and then press OK.
10. Confirm the adjustment by pressing Yes. (If you press
No, the adjustment mode display returns and no changes are
made.) The adjustment has now been completed.
11. To exit the adjustment mode, press Back.
12. To return to the basic display, press Exit.
76 __________________________________________________________________ M010091EN-F