Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 33
Set Serial Interface Mode
Use the SMODE command to set the default serial interface mode.
SMODE [xxxx]<cr>
where
xxxx
=
STOP, RUN or POLL
Table 8 Output Modes
Mode
Measurement Output
Available Commands
STOP
Only with the SEND command.
All (default mode).
RUN
Automatic output.
Only command S.
POLL
Only with the SEND [addr]
command.
SEND [addr] and OPEN [addr].
Other commands available after
opening a line to the transmitter
using the OPEN command.
Use with RS-485 buses where
multiple transmitters can share
the same line.
MODBUS
Must be read from the
transmitter using Modbus
protocol.
Modbus protocol only; see
Appendix A, Modbus Reference
on page 70.
When several transmitters are connected to the same line, each
transmitter must be entered an own address in the initial configuration,
and POLL mode must be used.
If the Modbus mode is activated, serial port commands cannot be used
with the device. To access the serial port command interface, follow the
procedure in Accessing Serial Port Command Interface from
Modbus Mode on page 44.