Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 61
5. When your serial connection is working, use the L command to see
the current user adjustment parameters. Write down the Cp offset
and gain values so you can restore them manually using the
LI
command, if necessary.
>l
6. Use the
ERRS command to see that no errors are active:
>errs
7. Install the HMK15 calibrator fitting over the LiCl salt chamber of
the humidity calibrator (11 %RH). Make sure it sits tightly over the
salt chamber.
8. Insert the DMT132 into the chamber. Do not remove the filter.
Make sure all unused holes are plugged and the transmitter is
installed tightly on the thread.
9. Give the
CRH command. The output from the command shows the
currently measured RH value, and prompts you to enter the value
of the first reference (Ref1). This is the reference on the dry end,
which produces the offset correction for the measurement.
>crh
RH: 11.39800112 Ref1 ?
10. Wait for 20 ... 40 minutes for the reading to stabilize. Press ENTER
without typing anything to update the value shown by the
CRH
command.
RH: 11.39800112 Ref1 ?
RH: 11.38377201 Ref1 ?
RH: 11.37988329 Ref1 ?
11. When the measurement shown by
CRH is stable, you can either
apply the correction or exit the procedure:
- To apply the correction, press ENTER one last time to update
the value, type 11 and press ENTER again (assuming you are
using the HMK15 and the LiCl salt).
- To exit the
CRH command without adjusting the measurement,
press the ESC key and exit this procedure.