Vaisala HMT360Series Betriebsanweisung Seite 47

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 97
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 46
Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 45
DIP Switch Functions
The table below is also printed on the protection board:
0505-285
Figure 23 DIP Switch Functions
1: Analog output test on/off
If you turn the switch to on position (up), you can force the outputs to states
4 mA, 12 mA and 20 mA by pressing buttons Up and Dn on the cover.
Outputs return to normal mode when switch is turned down.
2: Calibration Disabled/Enabled
The EEPROMs are write protected. If this switch is in the disabled position
(down), it does not allow any calibrations or scalings.
3 and 4: Calibration rh, t, analog
With these combinations you can perform relative humidity, temperature
or analog output calibrations with a multimeter or with the transmitter
display unit. Turn the DIP switches to the desired position according to the
table printed on the protective cover.
5: Output quantities
Determines whether the output units are metric (down) on non-metric.
NOTE
Keep this switch always in the disabled position during normal use of the
transmitter.
Seitenansicht 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 96 97

Kommentare zu diesen Handbüchern

Keine Kommentare