HMP260-U017en-1.5 Operating Manual
1997-11-25 53
8. TECHNICAL DATA
Specifications of the protection unit, see Appendix 9.
8.1 Relative humidity
Measuring range 0...100 %RH
Accuracy (including nonlinearity and repeatability)
maximum achievable accuracy when calibrated against high quality, certi-
fied humidity standards:
±1 %RH (0...90 %RH)
±2 %RH (90...100 %RH)
when calibrated against salt solutions (ASTM E104-85):
±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)
Response time (90 %) at +20 °C in 15 s
still air (with sintered filter)
Sensor HUMICAP
®
H-sensor (part no.
16663)
8.2 Temperature
Measuring range
HMP263 -40...+120 °C
HMP264 -40 ...+180 °C
HMP265 -40...+180 °C
Typical accuracy of electronics at ±0.1 °C
+20 °C (+68 °F)
Typical temperature dependence of 0.005 °C/°C
electronics
Sensor Pt 100 RTD 1/3 DIN 43760B
8.3 Calculated variables
Typical ranges
dewpoint temperature -40...+100 °C
mixing ratio 0...500 g/kg dry air
absolute humidity 0...600 g/m
3
wet bulb temperature 0...+100 °C