HMP260 SERIESTRANSMITTERSOperating ManualHMP260-U017en-1.55 January 1996 Vaisala 1996

Seite 2 - Back of cover page

HMP260 SERIESOperating Manual HMP260-U017en-1.54 1997-11-253. INSTALLATION3.1. Selecting the place of installationThe transmitters should be installed

Seite 3 - Contents

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module5 SPECIFICATIONSConnections Berg socketson the main board screw terminals 0.5

Seite 4

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop moduleINSTALLING AND USING THE DIGITAL CURRENT LOOP MODULEINSTALLING AND USING THE DIG

Seite 5

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.51 INSTALLATIONSwitch the transmitter off.Resistors R2, R3 and R4 between connect

Seite 6

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop module2 OPERATIONThe HMP260 transmitters can either be given an address or operated wi

Seite 7 - 1. PRODUCT DESCRIPTION

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.5The digital current loop module has separate lines for transmitting and receivin

Seite 8 - 2 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop module3 NETWORK CONFIGURATIONSingle loop operationBidirectional data on one pair and g

Seite 9 - 1997-11-25 3

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.5• Set the transmitter in POLL mode:>smode pollSerial mode : POLLNOTE 1The S

Seite 10 - 3. INSTALLATION

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop moduleDual loop wiring• Set the address of the transmitter, it can be any number betwe

Seite 11 - 1997-11-25 5

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.54 CHECKING THE SERIAL PORT NETWORK OPERATIONNormally measurement readings are as

Seite 12 - 6 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Appendix 4: Digital current loop modulePOLL modeIf a transmitter has been set to POLL mode, it will respond only to com

Seite 13 - 1997-11-25 7

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 53.2. MountingHMP263/264/265(EEx ia) IICSAFE AREA HAZARDOUS AREA EEx ia IIC T3/T4HMP263 s

Seite 14 - 8 1997-11-25

HMP260 SERIESAppendix 4: Digital current loop module HMP260-U017en-1.55 SPECIFICATIONSGalvanic isolation 1500 VAC/DC max. (1 min)Loop supply voltage 4

Seite 15 - 1997-11-25 9

HMP260 SERIESHMP260-U017en-1.5 Appendix 5: Error messagesERROR MESSAGESThe HMP260 transmitters go through a self-diagnostics procedure when the power

Seite 16 - 10 1997-11-25

HMP260 SERIESAppendix 5: Error messages HMP260-U017en-1.5Errors during operationThere are two types of errors that may occur during the operation of t

Seite 17 - SAFE AREA

HMP260 SERIESHMP260-U017en-1.5 Appendix 5: Error messagesThe LEDs indicate these errors as follows:¡¡¤¤ no frequency at all¡¡¤¡ a frequency missing fr

Seite 18

HMP260 SERIESHMP260-U017en-1.5 Appendix 6: Calculation formulas1 CALCULATIONThe HMP260 series transmitters measure relative humidity and temperature.

Seite 19 - Hazardous

HMP260 SERIESAppendix 6: Calculation formulas HMP260-U017en-1.52 CORRECTIONS AT HIGH PRESSUREThe saturation pressure of water vapour calculated with e

Seite 20 - 14 1997-11-25

BLUGRNYELBLKHMP 265 In automation system galvanically isolated inputsare recommended for current signalsMain board HMP230To control systemSupply volta

Seite 21 - 4. COMMISSIONING

HMP263, HMP264 AND HMP265INSTALLATIONToolQtypcsDrawnReviewApprDesignReplacesScaleidTitleArchnoSerial Sheetdoc noCooperator`sDwg noRevLtr QtyChange Rea

Seite 22 - OPENED COVER OF THE HMP260

HMP260 SERIESHMP260-U017en-1.5 Appendix 9: ApprovalsAPPROVALS AND SPECIFICATIONS OF THE PROTECTION UNITThe transmitter conforms to the European Standa

Seite 23

HMP260 SERIESOperating Manual HMP260-U017en-1.56 1997-11-25The transmitters should be mounted with the sensor head horizontally; thisway, any water co

Seite 24

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 7NOTEWhen using the support bar the lead-through piece andthe space between the duct wall an

Seite 25 - RX GND T

HMP260 SERIESOperating Manual HMP260-U017en-1.58 1997-11-253. Insert the body of the sensor head into the fitting body and screw thenut manually to th

Seite 26 - 20 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 9Fig. 3.4 Some examples on the installation of the HMP264 sensor head(boring)threaded sleeve

Seite 27 - 5. COMMANDS

HMP260 SERIESOperating Manual HMP260-U017en-1.510 1997-11-253.2.3. HMP265 transmitterto be sealedto be insulatedSAFE AREAHAZARDOUS AREAFig. 3.5 Instal

Seite 28

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 11a plugged holefor an Exi-approvedreference meterHAZARDOUS AREASAFE AREAFig. 3.7 Mounting t

Seite 29 - 1997-11-25 23

HMP260 SERIESOperating Manual HMP260-U017en-1.512 1997-11-25• push back the braid and press it between the two brass disks toachieve a full 360° groun

Seite 30

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 13NOTEWhen the cable is grounded as above, the metallic partsof the sensor head, the shield

Seite 31

This page intentionally left blank.Back of cover page.

Seite 32

HMP260 SERIESOperating Manual HMP260-U017en-1.514 1997-11-25VmA+-CURRENT/VOLTAGEOUTPUTSEXTERNALPOWER SUPPCH1+CH1-CH2+CH2-24 V +OPENED COVER OF THE HM

Seite 33

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 154. COMMISSIONINGWhen the HMP260 transmitters leave the factory, their measurement rangesan

Seite 34

HMP260 SERIESOperating Manual HMP260-U017en-1.516 1997-11-25CH2 CURRENT OUTPUTCH1 0 ... 1 VOLTAGE OUTPUTCH1CH2CH2CH1X15CURRENT OUTPUTS0 ... 20 / 4 ...

Seite 35

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 17Fig. 4.3 Spare jumpers4.3 Connecting the RS 232C serial busFig. 4.4 Serial bus connections

Seite 36

HMP260 SERIESOperating Manual HMP260-U017en-1.518 1997-11-25Fig. 4.5 Connection of cablesWhen the serial bus has been connected between the PC and the

Seite 37

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 19The RS 232C screw terminal cannot be used if an RS 485/422 serial moduleor a digital curre

Seite 38

HMP260 SERIESOperating Manual HMP260-U017en-1.520 1997-11-25jumper must be removed before the transmitter is reset. If the jumper is inplace when powe

Seite 39

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 215. COMMANDSAs the HMP260 transmitters are microprocessor based devices, theirconfiguration

Seite 40

HMP260 SERIESOperating Manual HMP260-U017en-1.522 1997-11-255.2 LED commandsNOTEIf the transmitter has a display/keypad cover, the LEDcommands cannot

Seite 41

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 235.3 Display/keypad commands5.3.1 Display modeIn the display mode the transmitters output m

Seite 42

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 iContents1. PRODUCT DESCRIPTION ...

Seite 43

HMP260 SERIESOperating Manual HMP260-U017en-1.524 1997-11-25numbers is ended with selecting a blank ' ' and pressing ENT. The last charac-te

Seite 44 - 6. CALIBRATION

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 25NOTEThe analogue output jumpers must be set to right places(see Fig. 4.2).5.3.4.2 Selectin

Seite 45 - RH 1 point cal. Change the

HMP260 SERIESOperating Manual HMP260-U017en-1.526 1997-11-255.3.5 Output via the serial bus5.3.5.1 Turning the serial interface echo ON/OFF• Select M

Seite 46

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 27Selecting stop bits:Full duplex/half duplex:The processor does not allow the following com

Seite 47

HMP260 SERIESOperating Manual HMP260-U017en-1.528 1997-11-255.3.5.4 Selecting the output units• Select Unit in the main menu:• Use the arrow keys to s

Seite 48

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 295.3.6.1 Setting the serial interface operation mode• Select Mode in the main menu; the fol

Seite 49

HMP260 SERIESOperating Manual HMP260-U017en-1.530 1997-11-25• Select More in the main menu, select More in the More menu andselect Mtim in the second

Seite 50

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 315.3.7.3 Setting the date• Select More in the main menu; select Date in the More menu:• If

Seite 51

HMP260 SERIESOperating Manual HMP260-U017en-1.532 1997-11-255.4 Serial commandsMore detailed descriptions of the serial commands can be found in Appen

Seite 52

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 335.4.1.2 Selecting and scaling the analogue output quantitiesDisconnect the security lock j

Seite 53

HMP260 SERIESOperating Manual HMP260-U017en-1.5ii 1997-11-255.4. Serial commands ...

Seite 54

HMP260 SERIESOperating Manual HMP260-U017en-1.534 1997-11-255.4.2 Output via the serial bus5.4.2.1 Starting the measurement outputR <cr>Starts o

Seite 55

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 355.4.2.4 Setting the output interval for the RUN modeINTV xxx yyy <cr>xxx = output

Seite 56 - 7. MAINTENANCE

HMP260 SERIESOperating Manual HMP260-U017en-1.536 1997-11-255.4.2.6 Selecting the output unitsUNIT x <cr>x = m(etric units)n(on-metric units)me

Seite 57

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 37In POLL mode: measurements output only with command SEND. When inPOLL mode, the output sta

Seite 58 - 52 1997-11-25

HMP260 SERIESOperating Manual HMP260-U017en-1.538 1997-11-256. CALIBRATIONThe HMP260 transmitters have been fully calibrated at the factory and theres

Seite 59 - 8. TECHNICAL DATA

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 39NOTEIf the sensor has been changed, the calibration has to bedone according to the instruc

Seite 60

HMP260 SERIESOperating Manual HMP260-U017en-1.540 1997-11-256.1.1.3 Using LED commandsDisconnect the security lock jumper!• Make sure that the sensors

Seite 61

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 41• If you want to see how the sensor stabilizes to the humidity in thecalibrator, enter c &

Seite 62 - 56 1997-11-25

HMP260 SERIESOperating Manual HMP260-U017en-1.542 1997-11-25• Wait for 10 minutes.• Connect an ammeter/voltmeter to the analogue outputs (connectorX2)

Seite 63 - 1997-11-25 57

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 43• Press any key and enter the second point value and press <cr>.RH : xx.x Ref2 ?

Seite 64 - 58 1997-11-25

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 iii8. TECHNICAL DATA...

Seite 65 - GUARANTEE

HMP260 SERIESOperating Manual HMP260-U017en-1.544 1997-11-25• Check that the reading corresponds within the desired accuracy to thereading in the cali

Seite 66

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 456.2.1 One point offset correction6.2.1.1 Using serial commandsDisconnect the security lock

Seite 67

HMP260 SERIESOperating Manual HMP260-U017en-1.546 1997-11-256.2.1.3 Using LED commandsDisconnect the security lock jumper!• Leave the reference instru

Seite 68 - 1 ANALOGUE OUTPUT COMMANDS

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 47• The stabilization of the sensor can be monitored by entering c <cr>instead of the

Seite 69

HMP260 SERIESOperating Manual HMP260-U017en-1.548 1997-11-256.3 Calibration of analogue outputsThe analogue outputs have been calibrated at the factor

Seite 70

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 49• Connect the meter to the output of channel 2 and enter the measuredlower end current/vol

Seite 71 - 2 CALIBRATION COMMANDS

HMP260 SERIESOperating Manual HMP260-U017en-1.550 1997-11-257. MAINTENANCE7.1 Reference measurementsReference measurements are necessary for the verif

Seite 72

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 51When errors have occurred, the transmitter outputs the error code (see Appen-dix 3 for all

Seite 73

HMP260 SERIESOperating Manual HMP260-U017en-1.552 1997-11-25CH1+CH1-TEST POINTSCH1+CH1-mARLX2Fig. 7.4 Circuit diagram of the analogue output current t

Seite 74

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 538. TECHNICAL DATASpecifications of the protection unit, see Appendix 9.8.1 Relative humidi

Seite 75 - 3 OUTPUT VIA THE SERIAL BUS

HMP260 SERIESOperating Manual HMP260-U017en-1.5iv 1997-11-25This page intentionally left blank.

Seite 76

HMP260 SERIESOperating Manual HMP260-U017en-1.554 1997-11-25Accuracies of the calculated quantities depend on the accuracies of the meas-ured quantiti

Seite 77

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 55Accuracy of wet bulb temperature (°C)Relative humidityTemp. 102030405060708090100-40 0.20

Seite 78

HMP260 SERIESOperating Manual HMP260-U017en-1.556 1997-11-25Connections screw terminals, 0.5 mm2 wires (AWG20), stranded wires recommendedOperating vo

Seite 79 - "xxx...xxx"

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 57Ø 13.5Ø 13.5cable length 2000,5000, 10 000 or 15 000 m2970Fig. 8.1 HMP263 sensor head dime

Seite 80 - 4800 E 7 1 FDX

HMP260 SERIESOperating Manual HMP260-U017en-1.558 1997-11-258.8 Electromagnetic compatibilityThe emission and immunity tests have been performed accor

Seite 81

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 599. OPTIONSMounting flanges for HMP265 aluminiumstainless steel for harsh conditionsInstall

Seite 82

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsSERIAL COMMANDS1 ANALOGUE OUTPUT COMMANDS ...

Seite 83

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5

Seite 84 - 4 OTHERS

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commands1 ANALOGUE OUTPUT COMMANDSAMODE Setting the analogue outputsDisconnect the security lock jum

Seite 85

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5ASEL RH T <cr>Ch1 (RH) lo 0.000 %RH ? <cr>Ch1 (RH) hi 100.000 %RH ? <cr

Seite 86

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 11. PRODUCT DESCRIPTIONNOTEBefore installing and using the HMP260 transmitterstudy carefully

Seite 87

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsThe output scales can also be given directly with the ASCL command.ASCL aaa.a bbb.b ccc.c d

Seite 88 - SAFETY SUMMARY

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.52 CALIBRATION COMMANDSCRH Relative humidity calibrationDisconnect the security lock jumper!C

Seite 89

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFCRH Relative humidity calibration after sensor changeDisconnect the security lock jumper!FC

Seite 90

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:CT <cr>

Seite 91

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsLI Entering linear correction coefficientsDisconnect the security lock jumper!LI <cr>

Seite 92

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.53 OUTPUT VIA THE SERIAL BUSR Starting the measurement outputR <cr>Starts output of mea

Seite 93

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsThe output mode can be changed with command FORM.DSEND Outputting readings of all connected

Seite 94

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5ECHO Turning the serial interface echo ON/OFFECHO xxx <cr>xxx = ON or OFFWhen the ech

Seite 95

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFORM Setting the output formatFORM <cr>"xxx...xxx"? zzz...zzz <cr>xx

Seite 96

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5An example of a format suitable for use in Microsoft Excel spreadsheets:FORM <cr>"

Seite 97

HMP260 SERIESOperating Manual HMP260-U017en-1.52 1997-11-25OptionsCalculation variables dewpoint temperature, mixing ratio, absolute hu-midity, wet bu

Seite 98

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsFDATE Adding date to outputFDATE xxx <cr>xxx = ON or OFFWhen FDATE is activated, the

Seite 99

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5The settings can be changed one parameter at a time or all parameters at once:SERI O H <c

Seite 100 - 5 SPECIFICATIONS

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsADDR Setting the transmitter addressADDR aa <cr>aa = address (0...99)The address is u

Seite 101

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5OPEN aa <cr>SMODE xxxx<cr>aa = address of the transmitterxxxx = STOP, RUN or P

Seite 102 - 1 INSTALLATION

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commands4 OTHERSITEST Testing the analogue outputsITEST <cr>orITEST a b <cr>a = current

Seite 103 - 2 OPERATION

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5MTIM <cr>Mtim : 4 ? 5 <cr>orMTIM 5 <cr>Mtim : 5PRES Setting the pressure

Seite 104

HMP260 SERIESHMP260-U017en-1.5 Appendix 1: Serial commandsor pressure is set to zero using XPRES. After this the pressure stored using commandPRES is

Seite 105 - 3 NETWORK CONFIGURATION

HMP260 SERIESAppendix 1: Serial commands HMP260-U017en-1.5VERS Name and version of the programmeVERS <cr>For example:VERS <cr>HMPS 260 / x

Seite 106 - RX GND TX

HMP260 SERIESHMP260-U017en-1.5 Appendix 2: Safety summarySAFETY SUMMARYIn this summary the following symbols are used:WARNINGA WARNING denotes a hazar

Seite 107

HMP260 SERIESAppendix 2: Safety summary HMP260-U017en-1.5CAUTIONOnly fuses with the required current rating and of thespecified type may be used. Do n

Seite 108

HMP260 SERIESHMP260-U017en-1.5 Operating Manual1997-11-25 32. TO BE NOTED WHEN MEASURING HUMIDITYIn the measurement of humidity and especially in cali

Seite 109 - RH= 24.4 %RH T= 29.1 'C

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port moduleINSTALLING AND USING THE RS 485/422 SERIAL PORT MODULE1 INSTALLATION...

Seite 110

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.51 INSTALLATIONSwitch the transmitter off.Resistors R2, R3 and R4 between conne

Seite 111 - ERROR MESSAGES

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module2 OPERATIONThe HMP260 transmitters can either be given an address or operated

Seite 112

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5The RS 485/422 module has separate lines for transmitting and receiving, butth

Seite 113

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module3 NETWORK CONFIGURATIONSingle loop operationBidirectional data on one pair is

Seite 114 - 1 CALCULATION

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5• Set the address of the transmitter; it can be any number between 1and 99. In

Seite 115

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module• Open the transmitter cover.• Pull out the RS 485/422 serial port module, if

Seite 116

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5NOTE 2The transmitter outputs no prompt (>) after the SMODEPOLL command and

Seite 117 - 94-07-04 KKe

HMP260 SERIESHMP260-U017en-1.5 Appendix 3: RS 485/422 serial port module4 CHECKING THE SERIAL PORT NETWORK OPERATIONNormally measurement readings are

Seite 118

HMP260 SERIESAppendix 3: RS 485/422 serial port module HMP260-U017en-1.5POLL modeIf a transmitter has been set to POLL mode, it will respond only to c