Technical Information
ACCURACIES OF THE VARIABLES
Accuracies of the calculated variables depend on the cali-
bration accuracy of the humidity and temperature sensors;
here the accuracies are given for ±2 %RH and ±0.2 °C.
Accuracies of wet bulb temperature °C (ambient pressure 1013 mbar)
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.20 0.20 0.20 0.20 0.20 0.20 0.20 0.20
-20 0.21 0.21 0.22 0.22 0.22 0.22 0.23 0.23
0 0.27 0.28 0.28 0.29 0.29 0.29 0.30 0.30 0.31 0.31
20 0.45 0.45 0.45 0.44 0.44 0.44 0.43 0.43 0.42 0.42
40 0.84 0.77 0.72 0.67 0.64 0.61 0.58 0.56 0.54 0.52
60 1.45 1.20 1.03 0.91 0.83 0.76 0.71 0.67 0.63 0.60
80 2.23 1.64 1.32 1.13 0.99 0.89 0.82 0.76 0.72 0.68
100 3.06 2.04 1.58 1.31 1.14 1.01 0.92 0.85 0.80 0.75
120 3.85 2.40 1.81 1.48 1.28 1.13 1.03 0.95 0.88 0.83
140 4.57 2.73 2.03 1.65 1.41 1.25 1.13 1.04 0.97 0.91
160 5.25 3.06 2.25 1.82 1.55 1.37 1.24 1.13 1.05 0.99
Accuracies of dewpoint temperature °C
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 1.86 1.03 0.76 0.63 0.55 0.50 0.46 0.43
-20 2.18 1.19 0.88 0.72 0.62 0.56 0.51 0.48
0 2.51 1.37 1.00 0.81 0.70 0.63 0.57 0.53 0.50 0.48
20 2.87 1.56 1.13 0.92 0.79 0.70 0.64 0.59 0.55 0.53
40 3.24 1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58
60 3.60 1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64
80 4.01 2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.70
100 4.42 2.41 1.74 1.40 1.19 1.05 0.95 0.87 0.81 0.76
120 4.86 2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84
140 5.31 2.91 2.10 1.69 1.44 1.26 1.14 1.05 0.97 0.91
160 5.80 3.18 2.30 1.85 1.57 1.38 1.24 1.14 1.06 0.99
Accuracies of mixing ratio g/kg (ambient pressure 1013 mbar)
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.003 0.003 0.003 0.003 0.003 0.004 0.004 0.004
-20 0.017 0.018 0.019 0.021 0.022 0.023 0.025 0.026
0 0.08 0.09 0.09 0.10 0.10 0.11 0.11 0.12 0.13 0.13
20 0.31 0.33 0.35 0.37 0.39 0.41 0.43 0.45 0.47 0.49
40 0.97 1.03 1.10 1.17 1.24 1.31 1.38 1.46 1.54 1.62
60 2.68 2.91 3.16 3.43 3.72 4.04 4.38 4.75 5.15 5.58
80 6.73 7.73 8.92 10.34 12.05 14.14 16.71 19.92 24.01 29.29
100 16.26 21.34 28.89 40.75 60.86 98.85 183.66 438.56
120 40.83 74.66 172.36 —
MEASURED VARIABLES
Relative humidity
Measurement range 0...100 %RH
Accuracy (including nonlinearity and repeatability)
when calibrated against salt solutions (ASTM E104-85):
±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)
maximum achievable accuracy when calibrated against
high-quality, certified humidity standards:
±1 %RH (0...90 %RH)
±2 %RH (90...100 %RH)
Response time (90 %) at +20 °C
in still air (with sintered filter) 15 s
Humidity sensor HUMICAP
®
sensor
Temperature
Measurement range
HMP231 -40...+60 °C
HMP233 -40...+80 °C or -40...+120 °C
HMP234, HMP235 & HMP238 -40...+180 °C
Typical accuracy of electronics at +20 °C ±0.1 °C
Typical temperature dependence ±0.005 °C/°C
of electronics
Temperature sensor Pt 100 IEC 751 1/3 class B
CALCULATED VARIABLES (DEPENDS ON MODEL TYPE)
Typical ranges
HMP231
dewpoint temperature -40...+60 °C
mixing ratio 0...160 g/kg d.a.
absolute humidity 0...160 g/m
3
wet bulb temperature 0...+60 °C
enthalpy -40...+460 kJ/kg
HMP233, HMP234, HMP235 & HMP238
dewpoint temperature -40...+100 °C
mixing ratio 0...500 g/kg d.a.
absolute humidity 0...600 g/m
3
wet bulb temperature 0...+100 °C
enthalpy -40...+1500 kJ/kg
Accuracies of absolute humidity g/m≈
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.004 0.004 0.005 0.005 0.005 0.006 0.006 0.006
-20 0.023 0.025 0.027 0.029 0.031 0.032 0.034 0.036
0 0.10 0.11 0.12 0.13 0.13 0.14 0.15 0.15 0.16 0.17
20 0.37 0.39 0.41 0.43 0.45 0.47 0.49 0.51 0.53 0.55
40 1.08 1.13 1.18 1.24 1.29 1.34 1.39 1.44 1.49 1.54
60 2.73 2.84 2.95 3.07 3.18 3.29 3.40 3.52 3.63 3.74
80 6.08 6.30 6.51 6.73 6.95 7.17 7.39 7.61 7.83 8.05
100 12.2 12.6 13.0 13.4 13.8 14.2 14.6 15.0 15.3 15.7
120 22.6 23.3 23.9 24.6 25.2 25.8 26.5 27.1 27.8 28.4
140 39.1 40.0 41.0 42.0 43.0 44.0 45.0 45.9 46.9 47.9
160 63.5 64.9 66.4 67.8 69.2 70.7 72.1 73.5 74.9 76.4
CONNECTORS
Vs = power supply
ao = analog outputs
Accuracies of enthalpy kJ/kg (ambient pressure 1013 mbar)
Relative humidity
Temp. 10 20 30 40 50 60 70 80 90 100
-40 0.21 0.21 0.21 0.21 0.21 0.21 0.21 0.21
-20 0.24 0.25 0.25 0.25 0.26 0.26 0.26 0.27
0 0.4 0.42 0.43 0.45 0.46 0.47 0.49 0.5 0.52 0.53
20 0.98 1.03 1.08 1.13 1.18 1.23 1.29 1.34 1.39 1.45
40 2.7 2.87 3.04 3.22 3.4 3.59 3.78 3.98 4.18 4.39
60 7.2 7.82 8.48 9.19 9.96 10.79 11.69 12.65 13.7 14.84
80 18.04 20.72 23.88 27.66 32.22 37.77 44.61 53.17 64.02 78.06
100 43.96 57.64 77.98 109.93 164.1 266.37 494.62 1180.46 —
120 111.6 203.93 470.52
24 POWER SUPPLY Analog Power Serial Alarm
outputs supply interface output
no connectors shared cable shared cable chared cable -
no connectors, alarm output unit shared cable shared cable chared cable cable
shared connector for Vs and ao shared connector - -
as above but with the alarm output unit shared connector - cable
connector for the RS232C shared cable shared cable connector -
2 connectors: RS232C and ao + Vs shared connector connector -
2 connectors: shared for RS485 and Vs - shared connector -
115/230 V POWER SUPPLY Analog Power Serial Alarm
outputs supply interface output
no connectors shared cable cable chared cable -
connector for ao connector cable - -