USER'S GUIDE____________________________________________________________________
136 _________________________________________________________________ M210704EN-E
Error States
In error state the quantity is not measured and the output is shown as
follows:
- analog channel outputs 0 mA or 0 V (you can use the serial line
command AERR or display/keypad to change this fault indication
value, see section Analog Output Fault Indication Setting on page
119)
- the serial port outputs stars (***)
- the cover LED is blinking
- optional display: error indicator is lit.
- The error indicator disappears when the error state is over and you
have checked the error message. Press the INFO button to display
the error message.
0706-017
Figure 73 Error Indicator and Error Message
Number refers to Figure 73 above:
1 = Error Indicator
You can also check the error message via the serial interface by using the
command ERRS. In case of constant error, please contact the Vaisala
Service Centers (contact information on page 139).