Temperature Setting
The ambient temperature value can be set by using the following
command:
MF_TEMP xxx<cr>
where: xxx = ambient temperature (C) x 10.
Example of setting desired temperature (25 C):
>MF_TEMP 250
MF_TEMP=250
>
Save the settings:
MAIN_SAVE F<cr>
Display Setting
The display operation mode, display accuracy and probe temperature can
be set by using the following command:
MF_DMODE x<cr>
where: x = 0 (zero)/1/2/3/7/8/9
0= no display
1= ppm-reading with one decimal
2= %-reading with two decimals
3=reading with no decimals
7=ppm-reading with no decimal
8= ppm-reading with 10 ppm accuracy
>mf_dmode 7
MF_DMODE=07 00 00 00 00
Save the settings:
MAIN_SAVE F<cr>