User's Guide ______________________________________________________________________
98 __________________________________________________________________ M210855EN-C
ADDR [aa]
Example: the barometer is configured to address 99.
>ADDR
Address : 2 ? 99
>
SEND
Use the SEND command to output the reading once in POLL mode:
SEND [aa]
SCOM
Use the SCOM command to show or set an alias for the SEND
command. The given alias cannot be a command already in use.
Example:
>scom meas
Send command : meas
>send
pressure = 1013.02 hPa
>measpressure = 1013.02 hPa
>
OPEN
When all barometers on the RS-485 bus are in POLL mode the OPEN
command sets one barometer temporarily to STOP mode so that other
commands can be entered.
where
aa = address (0 ... 255) (default = 0)
where
aa = address of the barometer
Seitenansicht 99