Vaisala MMT330 Bedienungsanleitung Seite 97

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 171
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 96
Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 97
Formatting Serial Line Message
NOTE
Instead of using the
FTIME and FDATE commands described in this
section, you can use the
FORM command with modifiers TIME and
DATE. See section
FORM on page 98.
FTIME and FDATE
FTIME and FDATE commands will enable/disable output of time and
date to the serial line. To add time to R and SEND outputs enter:
FTIME [x]<cr>
To add date to R and SEND outputs enter:
FDATE [x]<cr>
where
x
=
ON or OFF
Example:
>send
aw= 0.277 T= 23.5 'C
>ftime on
Form. time : ON
>send
00:16:07 aw= 0.277 T= 23.5 'C
>fdate on
Form. date : ON
>send
2000-01-01 00:16:15 aw= 0.277 T= 23.5 'C
>
General Settings
Changing Quantities and Units
To change quantities and units use serial commands or the optional
display/keypad. The MMT330 measures the following quantities:
- Water activity (aw)
- Temperature (T) (metric unit: ºC, non-metric unit:ºF)
NOTE
Only the quantities selected when ordering the device can be selected as
display output quantit
ies.
Seitenansicht 96