User’s Guide ______________________________________________________________________
112 __________________________________________________________________M210925EN-B
See also Chapter 10, Troubleshooting, on page 109. In case of constant
error, please contact Vaisala Service Center, see Vaisala Service
Centers on page 108.
Table 11 Error Messaging/Text Messages Table
Text message identifier (in
NMEA 0183 v3.0 protocol only)
Text Message Interpretation and action
01 Unable to measure error The requested parameters are
not activated in the message,
see Chapter 8, Sensor and Data
Message Settings, on page 95,
and check the parameter
selection fields.
02 Sync/address error The device address in the
beginning of the command is
invalid. Ask the device address
with the ?! (SDI-12) or ?<cr><lf>
(ASCII and NMEA) command
and retype the command with
the correct address.
03 Unknown cmd error The command is not supported,
use the correct command format,
see Chapter 7, Getting the Data
Messages, on page 61.
04 Profile reset Checksum error in configuration
settings during power-up.
Factory settings used instead.
05 Factory reset Checksum error in calibration
settings during power-up.
Factory settings used instead.
06 Version reset New software version in use.
07 Start-up Software reset. Program starts
from the beginning.
08 Use chksum xxx Given checksum not correct for
the command. Use the proposed
checksum.
09 Measurement reset The ongoing measurement of all
the sensors interrupted and
started from the beginning.
NOTE
When WMT52 is used on an RS-485 bus with other polled devices, the
error messaging feature shall always be disabled. This is done by the
following command: 0SU,S=N<crlf>.