PTB 200 1 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1PTB 200 DIGITAL BAROMETERS1 PRODUCT DESCRIPTION 22 INSTALLATIO
PTB 200 9 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12.7 Multiple transmitters on one RS 232C busIt is possible to
PTB 200 10 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.13 SERIAL COMMANDS3.1 GeneralWhen delivered from factory the b
PTB 200 11 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1The PTB 200 barometers have three sending modes: STOP, RUN an
PTB 200 12 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1Examples:>SERI<cr>1200 E 7 1>SERI 9600 E 7 1
PTB 200 13 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1The basic definition consists of the current and new format:&
PTB 200 14 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1UNIT Setting the pressure unitUNIT x<cr>wherex hPa*, kP
PTB 200 15 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1FILT Setting the filter parametersFILT xxx yyyy<cr>wh
PTB 200 16 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1SMODE Setting the sending modeSMODE xxxx<cr> STOP and R
PTB 200 17 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1ADDR Setting the barometer address (for POLL mode only)ADDR&l
PTB 200 18 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1SLEEP Setting and resetting the sleep modeSLEEP ON<cr>
PTB200 Digital BarometerAppendix 11994-08-16 1SERIAL COMMANDSFunction Command Commentserial bus settings SERIecho on/off ECHOoutput format FORMpressu
PTB 200 19 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.13.3 Operating commandsR Starting the measurement outputR<c
PTB 200 20 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1Example of outputting the current settings:>INTV<cr>
PTB 200 21 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1RESET Resetting the barometerRESET<cr>resets the barome
PTB 200 22 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.14 ADJUSTMENT AND CALIBRATIONThe PTB 200 barometers can be adj
PTB 200 23 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1The following pressure adjustments are possible:• offset adju
PTB 200 24 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1Example of an offset and gain adjustment:>LP<cr>Ref1
PTB 200 25 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1First precalibrate the barometer at up to eight pressure leve
PTB 200 26 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.15 TECHNICAL DATAOPERATING RANGEPressure range 600 ... 1100 hP
PTB 200 27 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1GENERAL CHARACTERISTICS (factory settings*)Supply voltage 10
PTB 200 28 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1Dimensions in mm (inches)27.5 (1.08)120 (4.72)65 (2.56)Ø 6.5
PTB 200 2 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.11 PRODUCT DESCRIPTIONThe PTB 200 A and PTB 201 A are fully com
PTB200 Digital BarometerAppendix 11994-08-16 1SERIAL COMMANDSFunction Command Commentserial bus settings SERIecho on/off ECHOoutput format FORMpressu
PTB 200 3 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12 INSTALLATION2.1 MountingThe dimensions and recommended mount
PTB 200 4 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12.2 GroundingA single electrical cable with a screen and five
PTB 200 5 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12.3 Power connectionsTwo wires, one for supply voltage and ano
PTB 200 6 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12.4 Serial output mode connectionsThe serial communication can
PTB 200 7 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.12.5 Pulse output mode connections (software version 1.05 or hi
PTB 200 8 (28)Digital BarometersSSD/Operating Manual 24th February 1993 PTB200-O0284-1.1The pulse frequency is about 4.5 kHz at baud rate 9600. The pu