Set Output Interval
Use the INTV command to set or display the output interval.
INTV [n xxx]<cr>
where
n
=
time interval = 0 ... 255
xxx
=
time unit = "S", "MIN", or "H"
Example:
intv 1 min
Output interval: 1 MIN
The shortest output interval (with n = 0) is approximately one second due
to the internal measurement cycle.
Output Reading Once
Use the SEND command to output the reading once in STOP mode.
The output format depends on which parameters the transmitter can
output.
SEND [aaa]<cr>
where
aaa
=
Transmitter address (0 ... 255). Must be specified if the
transmitter is in POLL mode, and a line has not been opened
using the OPEN command.
Example:
send
T= 25.2 'C aw= 0.299 H2O= 19 ppm RS= 29.9 %