USER'S GUIDE ______________________________________________________
38 _____________________________________________________ M010139EN-G
1. Connect the analog output signal cable connector to the indicator
base connector. Connect the screw terminal block as follows:
Brown wire: common wire (-)
Yellow-green wire: signal (+)
2. Open the MENU, press OPEN.
3. Select ►Functions, press .
4. Select ►Analog output, press .
5. Select 0.0 V to set the value for the 0.0V output signal, press SET
(If the analog output is ON, take it OFF). Set the low value by
using the arrow buttons. Press OK to make the setting.
6. Select 1.0V to set the value for the 1.0V output signal, press SET.
Set the high value by using the arrow buttons. Press +/- button to
choose the sign of the value. Press OK to make the setting.
7. Select Analog output on/off, press ON to activate the analog output
and return back to the basic display.
8. A wire picture sign appears on the upper left corner of the basic
display.
9. To deactivate the analog output function, go to MENU ►Functions
►Analog output ►Analog output on/off
and press OFF.