Chapter 7 __________________________________________________ Getting the Data Messages
VAISALA______________________________________________________________________ 105
Examples:
$WItXT,01,01,01,Unable to measure error*6D<cr><lf> (wind data
request when all the wind parameters were disabled from the wind
message).
$WITXT,01,01,03,Unknown cmd error*1F (unknown command
0XO!<cr><lf>).
$WITXT,01,01,08,Use chksum 2F*72 (wrong checksum used in MWV
query command)
WI = Talker identifier (WI = weather instrument)
TXT = Text transmission identifier.
xx = Total number of messages, 01 to 99
xx = Message number.
xx = Text identifier (see text message table)
c---c = Text message (see text message table)
* Checksum delimiter
hh = Two-character checksum for the query command.
<cr><lf> = Response terminator