
Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 85
where
xxx = STOP, RUN, POLL or SEND
Table 16 Selection of Output Modes
Mode Output Available Commands
STOP Only with the SEND command All (default mode)
RUN Automatic output Only command S
POLL Only with the SEND [addr]
command
Use with RS-485 buses, see
Operation of RS-485 Module on
page 107.
SEND One message at power-up.
Selected output mode will be activated after power outages.
INTV
Use the command INTV to set the outputting interval for the RUN
mode.
INTV [xxx yyy]
where
xxx
= Output inverval (0 ... 255). 0: the fastest possible output rate.
yy
= Unit (s, min or h)
Example:
>INTV 10 min
Output intrv. : 10 min
>
ECHO
Use the command ECHO to set the user port echo. The command
either enables or disables echo of characters received.
ECHO [x]
where
x = ON (default) or
= OFF
Kommentare zu diesen Handbüchern