User's Guide ______________________________________________________________________
18 __________________________________________________________________ M210669EN-D
Output with or without the Unit
.FORM.0<cr>
The command cancels the outputting of the unit after pressure reading.
This is also the factory setting for the output format.
1017.61
.FORM.1<cr>
The command adds the unit to the outputting. The standard output
field is for 6 characters. The first character is always a space.
1017.61 hPa
NOTE
To activate the new setting, please remember to reset the barometer
with the command .RESET<cr>
.RESET Resetting the Instrument
.RESET<cr>
The command resets the barometer and activates the setting changes.
Setting the Pressure Corrections
On/Off
These commands are used to set multi-point corrections either on or
off. See also the chapter Calibration and Adjustment on page 21.
.MPCON<cr>
The command activates the multi-point corrections.
.MPCOFF<cr>
The command deactivates the multi-point corrections.
NOTE
To activate the new setting, please remember to reset the barometer
with the command .
RESET<cr>