Appendix A ______________________________________________________________ Applications
VAISALA ________________________________________________________________________ 79
Technical Data
Typical measuring range
0...80* ppm (0...100ºC)
*Upper edge limited to saturation
Accuracy
(Calculation Model with
Average Coefficients)
Temperatures > 30 ºC: better than 10 % of
the reading
Temperatures < 30ºC: see Figure 23 below.
1311-206
Figure 23 Maximum Errors Caused by Deviation of Mineral Oils
Using Calculation Model with Average Coefficients
Temperature
Measurement range -40...+180ºC
Response times (with stainless steel filter)
In still air (20ºC) 10 s
In still transformer oil (20ºC) < 10 min