Chapter 10 _______________________________________Calibration and adjustment of transmitters
VAISALA _______________________________________________________________________ 35
Calibration and adjustment of HMP228 series
transmitters
You are able to check all the parameters a
w
, T and RH but only RH can be
adjusted. You can calibrate your transmitter against a calibrated reference
probe of MM70 or against a calibrator's reference humidity by using MI70
indicator in communication. Follow the first 12 steps and continue
according to the chosen calibration method.
1. Connect the 27159ZZ-calibration cable to the X5 connector of the
HMP228-series transmitter (X5 connector = 6-pin connector on the
left side of the motherboard).
2. Connect the other end of the calibration cable to either of the
MM70's connector port located on the bottom of the indicator, see
figure on page 34.
3. Turn on the MM70.
4. Open the MENU: press OPEN.
5. Select Functions, press .
6. Select Calibrate transmitters, press .
7. Select 1.Baud rate and 2.Serial format. Press SET to change the serial
settings. Serial settings of HMP228 and MM70 shall be the same.
(HMP228 factory serial settings: 4800, 7, even 1).
8. Select 3.Start calibration, press START. If the serial settings of MM70
are not compatible with the settings of HMP228, or the cable
connection is not working check that the serial settings of HMP228
and the settings of MM70 are the same. Check the cable connection.
Then retry connection in step 6.
9. After connection has been established start the
calibration/adjustment by pressing OK.
10. Select the quantity to check (a
w
/ T / RH) or to adjust (RH), press
SELECT.
11. Check the environment settings of MM70 probe, if needed. Press
YES or NO.
12. The first row shows the reading of the probe/transmitter connected to
the port I, the second row shows the reading of the probe/transmitter
connected to the port II. The third row ∆RH is the difference
between the readings of the ports.