Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 47
3. Remove the cap of the cable gland, including the rubber cylinder
and the metal rings. Slide the cap with rubber cylinder onto the
cable.
4. For a thin cable, add a shrinkable tube to increase cable
diameter.
5. Slide two metal rings on the shielding and squeeze it evenly
between the rings.
6. Secure with a shrinkable tube.
7. Tighten the cable with the cable gland and proceed with the
wiring.
8. Connect the signal cable to the screw terminals in the electronics
enclosure.
9. Ground the signal cable with the same method at both ends.
0205-006
Figure 14 Cable Grounding Instructions