APPENDIX 3 ________________________________________________________________________
114 ____________________________________________________________________ U263EN-1.4
>send 22
RH= 24.4 %RH T= 29.1 ’C
•
Disconnect the terminal.
• Check that the jumpers in connector X4 are in the right places.
X4
X4
X1 X2
• Remount the RS 485/422 serial module.
• Close the cover.
• Repeat this setting procedure with each transmitter.
• When all transmitters on the network have been configured,
switch them off.
Checking the serial port network operation
Normally measurement readings are asked when the transmitter is in
POLL mode; then the command must include the address of the
transmitter. If the settings of the transmitter need to be changed, the
transmitter is switched to STOP mode with command OPEN;
commands can then be given without address. When the line to the
transmitter is closed, it returns to POLL mode.
STOP mode
Open the line to the transmitter:
open 22<cr>
DMP 22 line opened for operator commands
Transmitter no. 22 is now temporarily set to STOP mode; it accepts
commands sent without address until CLOSE command is given.
Individual settings can now be easily modified. Do not open more
than one line at a time.