Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 35
List of Serial Commands
All commands can be issued either in uppercase or lowercase. In the
command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (Enter) key on
your computer keyboard. Enter a <cr> to clear the command buffer
before starting to enter commands.
Table 6 List of Serial Commands
Command
Description
?
Output information about the device
ADJD
View the adjustment date
AMODE
View or set the analog output mode
AOVER [ON/OFF]
Allow analog outputs to exceed their range
10%
ASEL
Set analog output parameters and scaling
CDATE
View the calibration date
CODE
View the order code of the probe
CRH
Calibrate and adjust RH measurement
CRHCLR
Clear adjustment of RH measurement
CT
Calibrate and adjust T measurement
CTCLR
Clear adjustment of T measurement
CTEXT
View the calibration information field
ERRS
List present probe errors
FILT [0.001 … 1]
Set the result filtering
FRESTORE
Restore factory settings
HELP
List available commands
INTV [0 ... 255 S/MIN/H]
Set the continuous output interval (for RUN
mode)
L
Displays user adjustment parameters
R
Start the continuous outputting
RESET
Reset the probe
S
Stop the continuous outputting
SDELAY [0255]
View or set serial line answer minimum
delay
SEND [0 ... 255]
Output the reading once
SNUM
View the serial number of the probe
UNIT
Select metric or non-metric output units
VERS
View software version of the probe